Surge and EFT/Burst Immunity: IEC 61000-4-5 and 61000-4-4

How IEC 61000-4-4 EFT/burst and IEC 61000-4-5 surge tests work: waveforms, coupling networks, typical levels, failure modes and protection circuits that pass.

Published: 2026-11-18 · Last updated: 2026-11-21

Next to ESD, two immunity tests appear on almost every product test plan: electrical fast transients (EFT, or burst) according to IEC 61000-4-4 and surges according to IEC 61000-4-5. Both inject transients into the cables of the equipment under test (EUT), but they model different events and break products in different ways. EFT is a train of small, very fast pulses that upsets logic; a surge is a single slow pulse with enough energy to destroy components.

What each test reproduces

EFT models the switching of inductive loads such as relays, contactors and motors: as the contacts open, the gap re-strikes repeatedly and bursts of nanosecond spikes travel along the installation wiring. A surge models much slower, high-energy transients from lightning near or on a line and from switching or faults in the power distribution system.

Each EFT pulse has a 5 ns rise time and a 50 ns duration (5/50 ns). The pulses come in bursts of 75: a 15 ms burst at a 5 kHz repetition rate or a 0.75 ms burst at 100 kHz, repeated every 300 ms. The surge generator, called a combination wave generator, produces a 1.2/50 µs open-circuit voltage and an 8/20 µs short-circuit current. The ratio of peak voltage to peak current, its effective output impedance, is 2 Ω, so a 1 kV setting drives up to 500 A into a short circuit.

ParameterEFT/burst (IEC 61000-4-4)Surge (IEC 61000-4-5)
Real-world sourceSwitching of inductive loads, relay and contactor arcingLightning-induced transients, power system switching and faults
Waveform5/50 ns pulses in 15 ms bursts (5 kHz) or 0.75 ms bursts (100 kHz), every 300 ms1.2/50 µs voltage, 8/20 µs current, single pulses
SpectrumUp to about 70 MHz (0.35 / 5 ns)Mostly below 100 kHz
Coupling to power linesCDN with 33 nF capacitors, common mode between the lines and the GRPCDN: 18 µF line-to-line (2 Ω), 10 Ω + 9 µF line-to-earth (12 Ω)
Coupling to signal linesCapacitive coupling clamp around the cableCoupling network with 40 Ω series resistance, through capacitors or gas arresters
ApplicationAt least 1 min per polarity at each coupling pointTypically 5 positive and 5 negative surges per coupling mode
Typical failureResets, lock-ups, corrupted data, glitching readingsBlown fuses, cracked varistors, damaged ICs, insulation flashover

Waveforms and coupling networks from IEC 61000-4-4 Ed. 3 and IEC 61000-4-5 Ed. 3.

How the tests are set up

The EFT test is done over a ground reference plane (GRP) connected to protective earth, with the EUT on a 0.1 m insulating support. The burst is applied in common mode: the coupling/decoupling network (CDN) injects it between the power conductors and the GRP, while its decoupling side keeps the pulses out of the mains supply. Signal and data cables pass through a capacitive coupling clamp, so the pulses reach them only through stray capacitance, just as they do when a cable runs next to contactor wiring in a cabinet.

For surge, the CDN couples line-to-line through 18 µF and line-to-earth through 10 Ω plus 9 µF. The extra 10 Ω represents the higher impedance of the earth path, so the peak current is lower: about 167 A at 2 kV line-to-earth, against 500 A at 1 kV line-to-line. On AC ports the surges are synchronized with the mains at 0°, 90°, 180° and 270°, because a surge that adds to the mains peak stresses the protection hardest. All levels below the specified one must also be tested. A scheme in which a fast TVS diode and a slower gas discharge tube (GDT) share the surge can fail at an intermediate level where neither device takes the current cleanly.

StandardEFT, AC powerEFT, signalSurge, AC line-to-lineSurge, AC line-to-earth
IEC 61000-6-1 (residential, commercial, light industry)±1 kV±0.5 kV±1 kV±2 kV
IEC 61000-6-2 (industrial)±2 kV±1 kV±1 kV±2 kV
CISPR 35 (multimedia equipment)±1 kV±0.5 kV±1 kV±2 kV
IEC 60601-1-2 Ed. 4 (medical)±2 kV±1 kV±1 kV±2 kV

The generic standards and CISPR 35 require criterion B. Signal port tests apply only to ports whose cables can exceed the length stated in each standard, and DC power ports have their own levels.

How products fail

EFT failures are rarely about damage. The burst returns to the GRP through stray capacitance, so common-mode current flows through the product wherever the impedance is lowest: through the PCB ground, across an isolation barrier, out along another cable. A reset during EFT usually means that this current crossed a reset line, a crystal or a long unfiltered trace to a microcontroller input. Because the pulses arrive as bursts of 75 every 300 ms, firmware without error handling can lock up even when a single pulse would do no harm.

Surge failures are mostly about energy and voltage. A varistor too small for the current cracks or degrades after a few surges; a protection device that clamps too high lets the bridge rectifier, bulk capacitor or switching transistor behind it see overvoltage; and insufficient clearance arcs between primary and secondary circuits or to the chassis. Line-to-earth surges also stress Y capacitors and the insulation of products without protective earth, and in a system they drive current along signal cables between two units earthed at different points.

Protection circuits that pass

Against EFT

The fast edge decides everything. With BW ≈ 0.35 / t_r, a 5 ns rise time corresponds to 70 MHz, and the knee frequency 0.5 / t_r is 100 MHz. At 70 MHz a 4.7 nF Y capacitor has a reactance of only 0.48 Ω, but 10 nH of lead and trace inductance in series adds 4.40 Ω. The connection, not the capacitor value, decides how well the filter works.

  • Filter at the point of entry: a common-mode choke with short-lead Y capacitors to chassis on the power input, and a ferrite bead or common-mode choke on I/O lines.
  • Give common-mode current a low-impedance path to chassis at the connector, so it does not travel across the PCB.
  • Filter slow inputs with RC networks. An ideal 1 kΩ / 10 nF filter has a cutoff of 15.9 kHz and about 36.0 dB attenuation at 1 MHz; capacitor ESL and coupling across the resistor reduce this at tens of megahertz, so place the capacitor at the IC pin with a short ground.
  • Keep reset, crystal and interrupt lines short and away from board edges and I/O cables, and filter the reset pin.
  • In firmware, use a watchdog, validate inputs, add CRC checks and retries on communication links, and refresh critical peripheral registers periodically.

Against surge

  • On AC mains, place a metal-oxide varistor (MOV) line-to-line after the fuse, and line-to-earth where the level requires it, often in series with a GDT to avoid leakage current and varistor ageing.
  • Choose the MOV continuous voltage above the highest mains voltage including tolerance, and its 8/20 µs current rating with margin for repeated surges.
  • Coordinate stages: a GDT or MOV takes most of the energy, and a series impedance (resistor, inductor or PTC) limits the current into a fast TVS diode that protects the sensitive circuit. Without the series impedance, the TVS diode clamps first and takes all the current.
  • Compare TVS pulse ratings on the same waveform. A part rated at 10/1000 µs handles a much higher peak current at 8/20 µs, and mixing the two leads to wrong choices.
  • Check that the fuse withstands the surge current without blowing, and check clearance and creepage across isolation barriers for the surge voltage, not only the working voltage.

Common mistakes

  • Testing only at the highest level and skipping the lower ones.
  • Using a small TVS diode for mains surge, or a high-capacitance MOV on a fast data line.
  • Connecting Y capacitors or TVS diodes through long traces, so the EFT edge sees inductance instead of a short circuit.
  • Running in-house EFT tests with a different support height or cable layout from the lab, which changes the stray capacitance.
Waveforms, coupling networks and levels in this article are taken from IEC 61000-4-4, IEC 61000-4-5 and the product standards listed. Always use the current edition of the standard that applies to your product. The 4.7 nF, 10 nH and 1 kΩ / 10 nF values are worked examples.